Sign In | Join Free | My ledscreensign.com
China China World Technology Medical Equipment Service Group logo
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group Win-win together!
Verified Supplier

5 Years

Home > Patient Monitor Parts >

6802-30-66765 Mindray T5 Patient Monitor Parts Integral Module Frame Rack

China World Technology Medical Equipment Service Group
Trust Seal
Verified Supplier
Credit Check
Supplier Assessment
Contact Now

6802-30-66765 Mindray T5 Patient Monitor Parts Integral Module Frame Rack

Brand Name : PHiliph

Model Number : 6802-30-66765

MOQ : 1pcs

Payment Terms : T/T(wire transfer), Western Union, Paypal

Delivery Time : 3-5 working days

Packaging Details : Carton

Product Name : 6802-30-66765 Mindray T5 Patient Monitor Integral Module Frame Rack

Brand : Mindray

Part No : 6802-30-66765

Condition : Used

Suitable For : T5

Warranty : 90 days

MOQ : One piece

Delivery Method : DHL,UPS ,Fedex ,etc

Packaging : Carton

Payment Method : T/T ,Paypal ,Western Union

Contact Now

6802-30-66765 Mindray T5 Patient Monitor Integral Module Frame Rack

Integral Module Rack

The patient monitors are available in two integrated modular racks: 2-slot and 5-slot. The control board contains a NIOS II FPGA. Realizes protocol conversion and infrared communication between host and parameter module

The module rack communication card can have 2 or 3 slots. The 3-slot communication card communicates directly with the motherboard. The 2-slot communication card is connected and controlled by the 3-slot communication card. The 3-slot communication card has a communication control function. The 2-slot communication card consists of an infrared circuit and a module power supply circuit. The RS422 driver board is located on the 3-slot communication board.

Removing the Integral Module Rack

  • Disconnect the cable that connects the integrated module frame to the motherboard. Unscrew the five M3 × 8 screws and remove the built-in module holder.

6802-30-66765 Mindray T5 Patient Monitor Parts Integral Module Frame Rack

  • Disconnect the cable that connects the communication board of the MPM frame to the communication board of the 3-slot module housing. Release the two buttons and remove the NiosII module from the 3-slot module holder.

6802-30-66765 Mindray T5 Patient Monitor Parts Integral Module Frame Rack

  • Loosen the eight M3 × 6 screws, and then remove the 3-slot modular rack communication card and the MPM modular rack communication card.

Note: Do not forget the ground spring when assembling.

6802-30-66765 Mindray T5 Patient Monitor Parts Integral Module Frame Rack

  • Loosen the hex nut with the sleeve. Then separate the disc, spring and contact pin.

6802-30-66765 Mindray T5 Patient Monitor Parts Integral Module Frame Rack

Module Rack Failures

Symptoms Possible Cause Corrective Action
Integral module rack cannot identify parameter modules Module failure Replace the parameter module. If a new module is identified, the original module is defective.
Cable defective or poorly connected

1. Check that the cables from the 3-slot module rack communication board to the MPM module rack communication board and from the module rack to the main board are properly connected.

2. Check that the connection cables and plugs are not damaged

Wrong communication board software revision Upgrade the program of the module or Integral module rack.
Module (in some slots) unrecognized Replace the corresponding module rack communication board.
Power supply to integral module rack abnormal 1. Check if voltage between two contact screws in any slot reaches 12VDC. If yes and the parameter module functions, PCB assembly in the SMR might fail. 2. If there is no 12V sent to the integrated module rack, check that power module output voltage to mother board reaches 12V DC. If yes, mother board might fail.
3-slot or MPM module rack communication board failure Replace the 3-slot or MPM module rack communication board.
Nios II module failure Replace the Nios II module.
Mother board failure Replace the mother board.

If you're interested in 6802-30-66765 Mindray T5 Patient Monitor Integral Module Frame Rack ,feel free to contact us .Thank you .


Product Tags:

6802-30-66765 Patient Monitor Parts

      

Integral Module Patient Monitor Parts

      
Quality 6802-30-66765 Mindray T5 Patient Monitor Parts Integral Module Frame Rack for sale

6802-30-66765 Mindray T5 Patient Monitor Parts Integral Module Frame Rack Images

Inquiry Cart 0
Send your message to this supplier
 
*From:
*To: China World Technology Medical Equipment Service Group
*Subject:
*Message:
Characters Remaining: (0/3000)